approx. 11,0 K/min., on average,
measured in the air inlet
Cooling-down rate
between 85 ... -40°C
approx. 9,5 K/min., on average,
measured in the air inlet
Heating-up rate as per
IEC 60068-3-5
approx. 10,0 K/min., on average,
measured in the air inlet
Heating-up rate
between -40 °C … +85 °C
approx. 10,0 K/min., on average,
measured in the air inlet
Temperature deviation, in time, in
steady state, depending on the temperature
±0,1 K to ±0,5 K
Temperature homogeneity, in space,
depending on the adjusted set point value3
±0,5 K to ±2,0 K
Heat compensation at -20 °C
at
+20 °C
8000 W
8000 W
Calibration
values
+23 °C and +80 °C
2) In case of continuous, long-term operation
below +5 °C ice might form on the evaporator if outer humidity is allowed to
enter the chamber, for instance, by open apertures, doors or if additional
humidity is generated by specimens or others.
In these cases, the evaporator has to be
defrosted from time to time by heating-up to 3+10 °C.
3) In the temperature range of minimum
temperature to +150°C.
approx. 11,0 K/min., on average,
measured in the air inlet
Cooling-down rate
between 85 ... -40°C
approx. 9,5 K/min., on average,
measured in the air inlet
Heating-up rate as per
IEC 60068-3-5
approx. 10,0 K/min., on average,
measured in the air inlet
Heating-up rate
between -40 °C … +85 °C
approx. 10,0 K/min., on average,
measured in the air inlet
Temperature deviation, in time, in
steady state, depending on the temperature
±0,1 K to ±0,5 K
Temperature homogeneity, in space,
depending on the adjusted set point value3
±0,5 K to ±2,0 K
Heat compensation at -20 °C
at
+20 °C
8000 W
8000 W
Calibration
values
+23 °C and +80 °C
2) In case of continuous, long-term operation
below +5 °C ice might form on the evaporator if outer humidity is allowed to
enter the chamber, for instance, by open apertures, doors or if additional
humidity is generated by specimens or others.
In these cases, the evaporator has to be
defrosted from time to time by heating-up to 3+10 °C.
3) In the temperature range of minimum
temperature to +150°C.
1.1.1 Performance Data for Temperature Tests
Maximum temperature
+180 °C
Minimum temperature
-72 °C, discontinuously 2)
Cooling-down rate as per
IEC 60068-3-5
approx. 11,0 K/min., on average, measured in the air inlet
Cooling-down rate
between 85 ... -40°C
approx. 9,5 K/min., on average, measured in the air inlet
Heating-up rate as per
IEC 60068-3-5
approx. 10,0 K/min., on average, measured in the air inlet
Heating-up rate
between -40 °C … +85 °C
approx. 10,0 K/min., on average, measured in the air inlet
Temperature deviation, in time, in steady state, depending on the temperature
±0,1 K to ±0,5 K
Temperature homogeneity, in space, depending on the adjusted set point value3
±0,5 K to ±2,0 K
Heat compensation at -20 °C
at +20 °C
8000 W
8000 W
Calibration values
+23 °C and +80 °C
2) In case of continuous, long-term operation below +5 °C ice might form on the evaporator if outer humidity is allowed to enter the chamber, for instance, by open apertures, doors or if additional humidity is generated by specimens or others.
In these cases, the evaporator has to be defrosted from time to time by heating-up to 3+10 °C.
3) In the temperature range of minimum temperature to +150°C.