This instrument is used to automatically sample gaseous samples into the gas analyzer. The number of sample points is highly customized, and up to 6 sample air bags can be added. The sample gas is transported by a vacuum pump from the air bag to the gas sample valve in the analyzer.
BAS-6型 6+1通道氣袋自動(dòng)進(jìn)樣器(6+1Sample Bag Auto Sampler)
一.應(yīng)用范圍
本儀器用于將氣態(tài)樣品批量自動(dòng)-進(jìn)樣至氣體分析儀。進(jìn)樣點(diǎn)位數(shù)量高度定制化,可加掛多6個(gè)采樣氣袋。樣品氣被一臺(tái)真空泵由氣袋輸送至分析儀中的氣體進(jìn)樣閥中。
This instrument is used to automatically sample gaseous samples into the gas analyzer. The number of sample points is highly customized, and up to 6 sample air bags can be added. The sample gas is transported by a vacuum pump from the air bag to the gas sample valve in the analyzer.
二.儀器特點(diǎn)
1.多位選擇閥,可進(jìn)行6位氣袋進(jìn)樣。
Multi-position valve, 6 positions.
2.能夠?qū)崿F(xiàn)對(duì)各品*儀進(jìn)行自動(dòng)進(jìn)樣控制,提高分析效率。支持的包括但不限于安捷倫、美國(guó)熱電、PE、島津等。
The auto sampler is compatible with multiple analyzer brands, including
Agilent, Thermo, PE, Shimadzu and etc.
3.具備彩色液晶顯示屏,能夠清楚的顯示儀器系統(tǒng)狀態(tài)
Instrument status clearly indicate at instument panel and analyzer, sampling or
idle.
4.進(jìn)樣口為標(biāo)準(zhǔn)1/8”穿板卡套接頭,更可根據(jù)用戶需求定制。
Standard 1/8” bulkhead connections, can be customized according to user
needs.
5.自動(dòng)吹掃,反吹系統(tǒng), 減少交叉污染
Auto system purge/ backflush function to prevent cross contamination.
6.使用真空隔膜泵進(jìn)樣,確保*的進(jìn)樣重現(xiàn)性。
Use diaphragm vacuum pump, ensure repeatability.
7.使用外徑為1/16”的樣品管路連接至分析儀的氣體進(jìn)樣閥。
1/16” tubing for connection to gas sample valve.
8.實(shí)現(xiàn)進(jìn)樣全自動(dòng)化,提高分析效率。
High automation, reduce manpower, improve efficiency, safety and etc.
9.可控制進(jìn)樣順序、重復(fù)次數(shù)、單次進(jìn)樣時(shí)長(zhǎng)、反吹時(shí)長(zhǎng),顯示當(dāng)前進(jìn)樣狀態(tài)等。
Analysis sequence, times of analysis, purge cycle time , compatible with
difference brand and model of analyzers
三.技術(shù)參數(shù)
項(xiàng)目
參數(shù)
產(chǎn)品貨號(hào)Product number
YXY01907
網(wǎng)絡(luò)連接Network Comm
RS485, 多同時(shí)可以控制4臺(tái)儀器, 大距離為1200米
進(jìn)樣口數(shù)量 Inlet
標(biāo)準(zhǔn)為6個(gè)
連接管路 Tubing
316SS OD1/16”
供電 Power supply
220-240VAC
50Hz- 2.3A
額定功率 Power con sumption
100W
大進(jìn)樣壓力 Max. Operating pressure
2 barg
允許樣品溫度 Sample temperature
+10 至 +40 °C
環(huán)境溫度 Ambient temperature
存儲(chǔ)溫度 Storage temperature
-15 至 +60 °C
防護(hù)等級(jí) Degree of protectoin
IP 20
外觀尺寸 Dimensions
W x H x D= 140 x 500 x 580mm
進(jìn)樣重復(fù)性Repeatability
RSD≤1%
實(shí)際進(jìn)樣流量 Flow rate
標(biāo)準(zhǔn)約為 300 ml/min*
控制 Control
液晶屏幕能夠顯示進(jìn)樣順序、重復(fù)次數(shù)、單次進(jìn)樣時(shí)長(zhǎng)、反吹時(shí)長(zhǎng)、當(dāng)前進(jìn)樣狀態(tài)等
四.應(yīng)用分析
氣袋自動(dòng)進(jìn)樣器用于將氣態(tài)樣品批量自動(dòng)- 進(jìn)樣至氣體 分析儀。
Bag auto-sampler is used to inject the sample into gas analyzer automatically.
五.系統(tǒng)需求
1.氣體分析儀應(yīng)具備氣體進(jìn)樣閥。
A gas sample valve on the analyzer interface.
2.分析儀器需具備外部連接端口。
The analytical instrument needs an external connection port.